VARIATIONS IN THE POWER OUTPUT OF THE PIEZON-MASTER-400 ULTRASONIC ENDODONTIC UNIT

Citation
M. Ahmad et al., VARIATIONS IN THE POWER OUTPUT OF THE PIEZON-MASTER-400 ULTRASONIC ENDODONTIC UNIT, International endodontic journal, 27(1), 1994, pp. 26-31
Citations number
NO
Categorie Soggetti
Dentistry,Oral Surgery & Medicine
ISSN journal
01432885
Volume
27
Issue
1
Year of publication
1994
Pages
26 - 31
Database
ISI
SICI code
0143-2885(1994)27:1<26:VITPOO>2.0.ZU;2-2
Abstract
The present study was undertaken to see if there was any variability i n the power output of Piezon-Master 400 ultrasonic files when driven u sing different generators, tranducers and file holders. The displaceme nt amplitude of the oscillating tip of the file in air was used as a m easure of the power output. The results showed that there was consider able variability in the power output of Piezon-Master 400 ultrasonic f iles of similar size and length when driven using different generators , transducers and file holders. In consideration of this, it is recomm ended that a calibration device be incorporated in the ultrasonic unit so that the operator will have some knowledge of when the unit is wor king at its maximum efficiency.