M. Ahmad et al., VARIATIONS IN THE POWER OUTPUT OF THE PIEZON-MASTER-400 ULTRASONIC ENDODONTIC UNIT, International endodontic journal, 27(1), 1994, pp. 26-31
The present study was undertaken to see if there was any variability i
n the power output of Piezon-Master 400 ultrasonic files when driven u
sing different generators, tranducers and file holders. The displaceme
nt amplitude of the oscillating tip of the file in air was used as a m
easure of the power output. The results showed that there was consider
able variability in the power output of Piezon-Master 400 ultrasonic f
iles of similar size and length when driven using different generators
, transducers and file holders. In consideration of this, it is recomm
ended that a calibration device be incorporated in the ultrasonic unit
so that the operator will have some knowledge of when the unit is wor
king at its maximum efficiency.