APPLICATION OF TOTALLY AUTOMATED ONLINE SAMPLE CLEAN UP SYSTEM FOR EXTRACTION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY SEPARATION OF PEPTIDE LEUKOTRIENES

Citation
I. Ramis et al., APPLICATION OF TOTALLY AUTOMATED ONLINE SAMPLE CLEAN UP SYSTEM FOR EXTRACTION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY SEPARATION OF PEPTIDE LEUKOTRIENES, Journal of pharmaceutical and biomedical analysis, 11(11-12), 1993, pp. 1135-1139
Citations number
9
Categorie Soggetti
Pharmacology & Pharmacy
ISSN journal
07317085
Volume
11
Issue
11-12
Year of publication
1993
Pages
1135 - 1139
Database
ISI
SICI code
0731-7085(1993)11:11-12<1135:AOTAOS>2.0.ZU;2-S
Abstract
We have developed a fully automated on-line extraction-purification me thod for peptide leukotrienes in biological fluids using a column-swit ching technique. Individual leukotrienes in HPLC-collected fractions a re determined by immunoassay techniques. Leukotrienes were extracted f rom nasal lavage samples and eluted from the solid-phase extraction ca rtridge into the HPLC column with methanol-ammonium acetate buffer (60 :40, v/v) pH 5.4, as a mobile phase. This method provides good recover ies, excellent resolution of leukotrienes and is suitable to be combin ed with off-line radioimmunoassay (RIA) or enzymoimmunoassay (EIA). Th e most important losses are observed after evaporation-concentration a nd lyophilization. We also have compared RIA with EIA determination. S ignificant differences are found between RIA values and EIA values.