Cg. Crandall et al., EVALUATION OF THE COSMED K2 PORTABLE TELEMETRIC OXYGEN-UPTAKE ANALYZER, Medicine and science in sports and exercise, 26(1), 1994, pp. 108-111
The purpose of this study was to determine the validity and evaluate t
he accuracy of a portable telemetric oxygen uptake analyzer (K2). Two
experiments were carried out: a) using a mechanical lung, the accuracy
of the K2 to measure oxygen fractions and minute ventilation followin
g 10 and 60 min of warm-up was determined; and b) two maximal graded e
xercise tests (GXT) on 15 subjects, one with the K2 system and the oth
er with a standardized breath-by-breath (BBB) system, while heart rate
(HR), minute ventilation (V-E) and oxygen uptake (VO2) were compared.
Following 10-min warm-up prior to calibration, the K2 underestimated
the true oxygen fraction as early as 5 min into the test, and this val
ue continued to decrease throughout the 30-min test. After 60 min of w
arm-up prior to calibration, the K2 accurately measured the true oxyge
n fraction for the first 15 min; at minute 20, and on to minute 30, th
e K2 underestimated the oxygen fraction. Ventilation volumes were not
affected by warm-up time. Minute ventilation during the K2 GXT was sig
nificantly higher than V-E for the BBB test. No significant difference
s were found between the HRs obtained with the BBB or K2 systems. No d
ifferences in VO2, for any stage of the GXT were identified between th
e K2 device, BBB device or when a respiratory exchange ratio (RER) cor
rection factor was applied to the K2 derived values. However, the RER
correction factor did minimize the VO2 differences between the BBB and
K2 systems. Therefore, we conclude that the K2 accurately measures VO
2 during a GXT however, its accuracy can be compromised by limitations
inherent to the system.