DUAL-BEAM SUBPICOSECOND BROAD-BAND INFRARED SPECTROMETER

Citation
Tp. Dougherty et Ej. Heilweil, DUAL-BEAM SUBPICOSECOND BROAD-BAND INFRARED SPECTROMETER, Optics letters, 19(2), 1994, pp. 129-131
Citations number
6
Categorie Soggetti
Optics
Journal title
ISSN journal
01469592
Volume
19
Issue
2
Year of publication
1994
Pages
129 - 131
Database
ISI
SICI code
0146-9592(1994)19:2<129:DSBIS>2.0.ZU;2-P
Abstract
A method for obtaining broadband multichannel IR spectra with subpicos econd time resolution is described. Spectrally broad IR pulses, produc ed by difference-frequency mixing of two visible pulses in LiIO3, are split to make signal and reference pulses. After sample absorption, th e IR pulses are upconverted by a second LiIO3 crystal to produce visib le pulses that are dispersed on a CCD detector to yield transient spec tra with 100-cm(-1) bandwidth and 4-cm(-1) resolution. With this dual- beam technique, spectra with sensitivities of 0.001 Delta OD are obtai ned.