GLUTATHIONE AMPEROMETRIC ENZYME MICROSENSOR

Citation
Zx. Huang et al., GLUTATHIONE AMPEROMETRIC ENZYME MICROSENSOR, Analytical letters, 27(2), 1994, pp. 263-271
Citations number
19
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032719
Volume
27
Issue
2
Year of publication
1994
Pages
263 - 271
Database
ISI
SICI code
0003-2719(1994)27:2<263:GAEM>2.0.ZU;2-3
Abstract
Glutathione micro-enzyme sensors were developed based on the immobiliz ation of glutathione oxidase at the tip of a 25 mum Pt wire sealed in glass. An inner membrane constructed from cellulose acetate (CA) was d eposited onto the tip of the Pt microelectrode prior to enzyme immobil ization with glutaraldehyde. The final outer diameter of the microelec trodes is approximately 30 mum. The analytical characteristics of the microelectrodes, including calibration curves, apparent K(M)', pH resp onse curves, stability and selectivity over enzymatic interferences we re determined. The response was linear in the concentration range 1.0 x 10(-5) M - 1.9 x 10(4) M glutathione, reaching 95% steady-state curr ent in 15-20 seconds. The microelectrodes were useful for more than tw o months.