A PERFORMANCE EVALUATION OF BARE ALPHA TRACK DETECTORS

Citation
Wm. Yeager et al., A PERFORMANCE EVALUATION OF BARE ALPHA TRACK DETECTORS, Health physics, 66(2), 1994, pp. 163-171
Citations number
18
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00179078
Volume
66
Issue
2
Year of publication
1994
Pages
163 - 171
Database
ISI
SICI code
0017-9078(1994)66:2<163:APEOBA>2.0.ZU;2-B
Abstract
Bare alpha track detectors are sensitive to radon progeny as well as t o radon gas. This paper reports the test results for two brands of bar e alpha track detectors that received several radon exposures at three progeny concentrations. The results show a relationship that is a lin ear combination of both the radon gas and the radon progeny concentrat ions. The reported radon concentration depended on the equilibrium fac tor assumed by the processor. The sensitivity of bare alpha track dete ctors increases with altitude, but this is a minor, correctable effect compared to the equilibrium factor.