Purpose: To determine the effective time window for MR imaging of tumo
rs with blood-brain barrier (BBB) damage after injection of Gd-contain
ing contrast media. Material and Methods: Eleven patients with 15 brai
n lesions (metastasis, glioma, abscess) were studied with a T1-weighte
d spin-echo sequence repeated over periods of up to 43 min after contr
ast injection. A quotient was calculated between the signals in the le
sion and in the gray matter, and plotted against time. Results: All le
sions reached 70% of the maximum RATIO within 3.5 min. After 25 min 12
out of 15 lesions showed persistent enhancement within 15% of the max
imal RATIO. Conclusion: The peak enhancement of BBB damage occurs arou
nd 3.5 min after injection. The effect does not change during the next
25 min. Scanning should not be started until 2-5 min after injection
of the contrast medium. and there is no advantage in waiting longer as
no major increase (or decline) of contrast can be expected.