IMPROVING THE PERFORMANCE OF THE COMBINATION PERSONAE NEUTRON DOSIMETER

Citation
Ma. Buckner et al., IMPROVING THE PERFORMANCE OF THE COMBINATION PERSONAE NEUTRON DOSIMETER, Radiation protection dosimetry, 51(1), 1994, pp. 5-15
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
51
Issue
1
Year of publication
1994
Pages
5 - 15
Database
ISI
SICI code
0144-8420(1994)51:1<5:ITPOTC>2.0.ZU;2-7
Abstract
The combination personal neutron dosemeter utilises two bubble detecto rs and a four element albedo thermoluminescence dosemeter to acquire a simple four interval neutron spectrum. Appropriate conversion factors may be applied to the measured fluence values to determine pertinent dosimetric quantities. The spectrometric and dosimetric accuracy of th e original design has been improved by refining the components' respon se functions and by applying a matrix algorithm for spectrum deconvolu tion. The modifications yielded a net improvement in dosimetric accura cy of 28% for the five radioisotopic spectra and 2% for the five in si tu spectra that were evaluated. The modifications and an assessment of their effectiveness in improving neutron measuring capabilities are d iscussed.