M. Kobayashi et al., YALO(3)CE-AM LIGHT PULSERS AS A GAIN MONITOR FOR UNDOPED CSI DETECTORS IN A MAGNETIC-FIELD, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 337(2-3), 1994, pp. 355-361
YAlO3-Am pulsers show no degradation in their light output or resoluti
on upon the application of a magnetic field, at least up to 10 kG. The
decrease in the pulse height agrees with the change in the PMT gain c
alibrated by an LED. When the pulser is mounted on the top face of a 3
0 cm long CsI far from a PMT with a UV filter inserted between the PMT
and the CsI, the light output decreases by a factor of 1/4, compared
with the case of being directly mounted on the PMT window. In the abov
e configuration, the 5.49 MeV a of Am-241 gives about 550 photoelectro
ns, which corresponds to a gamma-equivalent energy of 6.6 MeV. Finally
, the pulsers were mass-produced with a FWHM resolution of 5.4% (avera
ge) for the pulse height.