EVALUATION OF THE RESPONSE MATRIX OF A BONNER SPHERE SPECTROMETER WITH LII DETECTOR FROM THERMAL-ENERGY TO 100 MEV

Citation
V. Mares et H. Schraube, EVALUATION OF THE RESPONSE MATRIX OF A BONNER SPHERE SPECTROMETER WITH LII DETECTOR FROM THERMAL-ENERGY TO 100 MEV, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 337(2-3), 1994, pp. 461-473
Citations number
26
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
337
Issue
2-3
Year of publication
1994
Pages
461 - 473
Database
ISI
SICI code
0168-9002(1994)337:2-3<461:EOTRMO>2.0.ZU;2-I
Abstract
The neutron responses of a Bonner sphere spectrometric system with a 4 mmx4 mm (LiI)-Li-6 scintillator have been calculated using the MCNP M onte Carlo code in the energy range from thermal to 100 MeV. Calculate d responses have been interpolated to evaluate the response matrix wit h 51 energy points and 33 sphere diameters. Responses have been parame trized as a function of sphere mass by a hybrid log-normal function. F inally, the new response matrix has been compared with others reported in the literature and with recent experimental findings.