Pg. Kerkar et al., UNUSUAL TEAR IN INOUE BALLOON DURING PERCUTANEOUS BALLOON MITRAL VALVULOPLASTY IN A PATIENT WITH CALCIFIC MITRAL-STENOSIS, Catheterization and cardiovascular diagnosis, 31(2), 1994, pp. 127-129
An unusual tear in an Inoue balloon during dilatation of calcific mitr
al stenosis is presented and its mechanisms discussed. An abnormal seq
uence of inflation indicates a possible tear. (C) 1994 Wiley-Liss, Inc
.