Nf. Schrage et al., SIMPLE EMPIRICAL CALIBRATION OF ENERGY-DISPERSIVE X-RAY-ANALYSIS (EDXA) ON THE CORNEA, Scanning microscopy, 7(3), 1993, pp. 881-888
Monitoring of the corneal electrolyte content is important for the stu
dy of chemical eye burns. This paper describes quantitative measuremen
ts on gelatin standards, corneas and a cornea homogenate with an energ
y dispersive X-ray analyzer (EDX) in the scanning electron microscope
(SEM). Ten micrometers thick cryosections were freeze-dried and mounte
d on solid carbon supports. The applied quantification procedure was a
local peak background analysis with a specifically designed computer
program. Similar chemical and physical properties of gelatin, cornea h
omogenate, and cornea were proven by EDX-analysis and wet chemical ana
lysis. Gelatin standards with known concentrations of different added
salts showed linear correlations with a correlation coefficient higher
than 0.95 for all considered elements. The local background generatio
n on carbon supports was the same for gelatin standards and corneal ti
ssue. The results demonstrate that quantitative EDX analysis of semi-t
hin samples, mounted on neutral carbon supports, can be reliably used
for the assessment of the corneal mineral composition.