A survey is given of the actual possibilities and limitations of the u
se of linear accelerators (Linac radiosurgery systems) for intra = cra
nial radiosurgery. Depending on the collimator size, spherical fields
from 5-54 mm in diameter can be irradiated with dose gradients from 10
% (large fields) to 20% (small fields) per millimeter distance between
surface and treatment volume. This is comparable to the possibilities
of Gamma-Knife and Proton-irradiation. Optimal mechanical adjustment
of gantry and linac table are necessary for the required stability of
the isocenter. Mechanical inaccuracy should be smaller than 0.8 mm. Ad
vanced computerized 3D-treatment planning systems are indispensable pr
erequisites for accurate treatment and use of the flexibility of the l
inac system. Future developements are outlined.