ELECTROCHEMICAL CHARACTERISTICS OF H2O2 MICROARRAY ELECTRODES AS BASEELEMENTS OF BIOSENSORS

Citation
Jz. Zhu et al., ELECTROCHEMICAL CHARACTERISTICS OF H2O2 MICROARRAY ELECTRODES AS BASEELEMENTS OF BIOSENSORS, Fresenius' journal of analytical chemistry, 348(4), 1994, pp. 277-280
Citations number
12
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
348
Issue
4
Year of publication
1994
Pages
277 - 280
Database
ISI
SICI code
0937-0633(1994)348:4<277:ECOHME>2.0.ZU;2-J
Abstract
The electrochemical characteristics were examined of band- and square- type MAEs (microarray electrodes) with 10, 20 and 30 mu m band width a nd side length prepared by using standard planar processing in Kg Fe(C N)6 solution. Among them, S1-5 square-type MAE with 10 mu m side lengt h and 9.0 x 10(-3) mm(2) in the total area exhibits the largest curren t density and the shortest response time. The H2O2 calibration curve o btained at the S1-5 shows linearity from 0.1 to 5.0 mmol/l, a mean slo pe of 550 nA/mmol/l with a CV (variation coefficient) of 16.1%. A gluc ose sensor based on the S1-5 was prepared and its sensitivity was 21nA /mmol/l, ten-fold greater than that of a single ME (microelectrode) re ported lately.