A. Uritani et al., ANALYTICAL CALCULATIONS ON RADIOACTIVITY MEASUREMENTS OF IR-192 METALLIC SOURCES WITH A CALIBRATED IONIZATION-CHAMBER, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 339(1-2), 1994, pp. 377-380
The self-attenuation of photons was analytically calculated for an Ir-
192 metallic wire source and an aqueous standard source. The influence
of the self-attenuation on the current of the calibrated ionization c
hamber was also evaluated. The self-attenuation did considerably affec
t the ionization current. It was concluded that the apparent radioacti
vity must be multiplied by a correction factor larger than unity (1.09
+/-0.01 in the present case) to get the true radioactivity when the ra
dioactivity of the Ir-192 metallic wire source was measured with the i
onization chamber calibrated with the aqueous standard source of Ir-19
2.