R. Borio et al., OPTIMIZATION NEED OF DENTAL RADIODIAGNOSTIC PROCEDURES - RESULTS OF EFFECTIVE DOSE EVALUATION FROM RANDO PHANTOM MEASUREMENTS, Radiation protection dosimetry, 51(2), 1994, pp. 137-140
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
Radiological examinations of different types are needed in dental prac
tice both to make a correct diagnosis and to carry out, consequently,
an adequate therapy. Particularly in orthodontic practices. because of
the youth of the majority of the patients, an assessment of the detri
ment to health (through the effective dose evaluation) caused by medic
al diagnostic exposure to ionising radiation is needed to make decisio
ns about the optimisation of dental radiodiagnostic procedures. Experi
mental data from measurements on a Rando phantom were collected for th
e radiological examinations required for dental and for orthodontic pr
actices (with and without protective apron and collar). The results sh
ow the effectiveness of the leaded apron and collar in all the examina
tions carried out, particularly in reducing thyroid dose.