MEASUREMENT OF DEAD-SPACE IN A MODEL LUNG USING AN OSCILLATING INSPIRED ARGON SIGNAL

Citation
Em. Williams et al., MEASUREMENT OF DEAD-SPACE IN A MODEL LUNG USING AN OSCILLATING INSPIRED ARGON SIGNAL, Acta anaesthesiologica Scandinavica, 38(2), 1994, pp. 126-129
Citations number
12
Categorie Soggetti
Anesthesiology
ISSN journal
00015172
Volume
38
Issue
2
Year of publication
1994
Pages
126 - 129
Database
ISI
SICI code
0001-5172(1994)38:2<126:MODIAM>2.0.ZU;2-1
Abstract
In a model lung, airways dead-space can be accurately measured using a forced inspired oscillating argon signal, which varies sinusoidally a bout a mean concentration of 6% v/v with an amplitude of +/-4% v/v. Wi th sinusoid forcing periods longer than 120 seconds, and at a breathin g rate of 13.4 breaths minute(-1), the mean airways dead-space can be measured with a standard error of less than 5%. Sinusoid forcing perio ds shorter than 120 s provided inaccurate estimates of dead-space and so should not. be used with this technique.