OPTIMIZATION OF METAL PHOSPHOR SCREENS FOR ONLINE PORTAL IMAGING/

Citation
B. Wowk et al., OPTIMIZATION OF METAL PHOSPHOR SCREENS FOR ONLINE PORTAL IMAGING/, Medical physics, 21(2), 1994, pp. 227-235
Citations number
23
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00942405
Volume
21
Issue
2
Year of publication
1994
Pages
227 - 235
Database
ISI
SICI code
0094-2405(1994)21:2<227:OOMPSF>2.0.ZU;2-N
Abstract
Studies were conducted to determine the optimal metal/phosphor screen for on-line video verification of radiation treatment portals. Screens were evaluated for luminance and spatial resolution as a function of composition and thickness at 6- and 23-MV x-ray energies. A new video technique was used to determine modulation transfer functions. Gd2O2S was found to be the most efficient (brightest), phosphor for this appl ication. Luminance was found to vary linearly with phosphor thickness up to a thickness of 500 mg/cm(2). Metal plates made of iron, brass, c opper, lead, and sintered tungsten of various thicknesses were also te sted for luminance and resolution with Gd2O2S phosphor. Brightness pea ked at about 2-mm thickness for most metals. Significant contributions to the brightness were found to come from x rays interacting with the phosphor itself.