V. Kretschmer et al., GERMAN MULTICENTER STUDIES ON PLATELETPHERESIS FROM HEALTHY DONORS, International journal of artificial organs, 16, 1993, pp. 125-129
It is reported on two German Multicenter Plateletpheresis Studies with
the Fresenius AS 104 cell separator accompanied by a Multicenter Coun
ting Study. 15 centers participated, a total of 1884 separations were
evaluated. It could be shown that multicenter studies help to give a m
ore objective opinion about a cell separator and the separation protoc
ol used if it is possible to improve the counting methods and to elimi
nate individual center effects by an accompanying Multicenter Counting
Study. The AS 104 proved to be a good cell separator for platelet col
lection regarding platelet yields (3.1 x 11(11) from 3.5 \ citrated bl
ood), separation efficiency (about 45%), yield prediction (SD =double
dagger 15%) and leukocyte contamination (median WBC 47x10(6)).