Xa. Li et al., CHARACTERISTICS OF ORTHOVOLTAGE X-RAY THERAPY BEAMS AT EXTENDED SSD FOR APPLICATORS WITH END-PLATES, Physics in medicine and biology, 42(2), 1997, pp. 357-370
The effects of extended SSD (ESSD) on orthovoltage x-ray therapy dosim
etry have been studied for a Therapax DXT300 treatment unit with sever
al open- and close-ended cones. For the open-ended cones, the beam out
put and percentage depth dose (PDD) at ESSD can be accurately estimate
d from the measured data at the nominal SSD (NSSD) by an inverse squar
e law (ISL) and a divergence factor for gaps (distance from cone end t
o point of interest) up to 15 cm. In the case of close-ended applicato
rs (with an end plate of 3.2 mm thick PMMA), extending SSD results in
significant changes in output and in PDDs, but has little effect on be
am penumbra and flatness for gaps up to 15 cm. The output at ESSD calc
ulated by applying the ISL with the NSSD is found to be as much as 9%
different from the measurement for the close-ended cone. The output at
ESSD can, however, be calculated accurately by applying the ISL with
an effective SSD or using a measured gap factor. The change in the PDD
s with SSD can be calculated from a geometric divergence using a pseud
o SSD which is approximately equal to 60% of the effective SSD determi
ned in air. The calculated PDD curves at ESSD using this divergence fa
ctor agree well with measurements (within +/-1% for beams of 100-300 k
Vp with cone size ranging from 10x10 to 20x20 cm(2) and gap distance u
p to 15 cm).