INTERCOMPARISON OF CRYOGENIC RADIOMETERS USING SILICON TRAP DETECTORS

Citation
A. Lassila et al., INTERCOMPARISON OF CRYOGENIC RADIOMETERS USING SILICON TRAP DETECTORS, Measurement science & technology, 8(2), 1997, pp. 123-127
Citations number
13
Categorie Soggetti
Instument & Instrumentation",Engineering
ISSN journal
09570233
Volume
8
Issue
2
Year of publication
1997
Pages
123 - 127
Database
ISI
SICI code
0957-0233(1997)8:2<123:IOCRUS>2.0.ZU;2-P
Abstract
An intercomparison of cryogenic radiometers of HUT/VTT (Finland), PTB (Germany) and SP (Sweden) using silicon trap detectors was performed. Also results of monitoring of the responsivities of trap detectors ove r a 6 year period are presented. It was noticed that several non-ideal ities can undermine the agreement between results of calibrations. The non-uniformity of the responsivity of trap detectors made the results sensitive to alignment and beam size, The nonlinearity of some trap d etectors was found to be large. The average relative differences in th e results of the laboratories, when all known corrections are applied, are +1 x 10(-4) for HUT-PTB, -3.7 x 10(-4) for SP-PTB and -4.3 x 10(- 4) for SP-HUT,with standard uncertainties of 4.3 x 10(-4), 3.0 x 10(-4 ) and 4.4 x 10(-4), respectively.