We have measured oxygen uptake (VO2) and carbon dioxide production und
er steady-state conditions in 14 conscious volunteer subjects breathin
g from the Mapleson A system at different fresh gas flow rates, by col
lecting gas vented from the breathing system. Mixed vented gas concent
ration was measured wish a mass spectrometer and volume with a water s
pirometer. Gas exchange and total ventilation were measured wish subje
cts under resting conditions breathing room air, and then whilst breat
hing from a Magill system. Mean values of VO2, at mean fresh gas flow
rates 1.43, 0.88, 0.67 and 0.49 times the resting total ventilation of
the subject, were 3.30, 3.30, 3.34 and 3.56 ml min(-1) kg(-1) (STPD),
respectively. Rebreathing occurred at ratios of 0.67 and 0.49. We wer
e unable to demonstrate any increase in VO2, in the Mapleson A system
with rebreathing.