CRITICAL ANALYSIS OF THE ERROR IN THE EFFICIENCY MEASUREMENT OF GERMANIUM DETECTOR

Citation
S. Mukherjee et Jr. Rao, CRITICAL ANALYSIS OF THE ERROR IN THE EFFICIENCY MEASUREMENT OF GERMANIUM DETECTOR, National Academy Science Letters, 16(3), 1993, pp. 107-109
Citations number
NO
Categorie Soggetti
Multidisciplinary Sciences
ISSN journal
0250541X
Volume
16
Issue
3
Year of publication
1993
Pages
107 - 109
Database
ISI
SICI code
0250-541X(1993)16:3<107:CAOTEI>2.0.ZU;2-8
Abstract
It is convenient to use a multigamma ray source like Eu-152 for the ca libration of germanium detectors and to make a straight line fitting o f the detection efficiency vs. photon energy on a log-log representati on for the ease in interpolation. Considerable uncertainties are assoc iated with such interpolation as well as the error analysis, particula rly in the energy regions 100-200 KeV and 500-700 KeV, where the depar tures from a straight line are most significant. A more realistic appr oach would be to make a linear plot of a the ratio, epsilon0/epsilon(f ) vs. photon energy, where epsilon(0) and epsilon(f) are the measured and fitted efficiencies for prominent gamma rays of Eu-152. The result ing curve, which magnifies the departures from straight line, allows a more accurate interpolation and error analysis.