CALIBRATION OF LOW ACTIVITY CESIUM TUBES AND NEEDLES TRACEABLE TO THETHERAPY LEVEL STANDARD

Citation
Na. Deshpande et Jm. Wilkinson, CALIBRATION OF LOW ACTIVITY CESIUM TUBES AND NEEDLES TRACEABLE TO THETHERAPY LEVEL STANDARD, British journal of radiology, 67(794), 1994, pp. 194-199
Citations number
11
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
British journal of radiology
ISSN journal
00071285 → ACNP
Volume
67
Issue
794
Year of publication
1994
Pages
194 - 199
Database
ISI
SICI code
Abstract
A technique for deriving the reference air kerma rate for low activity brachytherapy sources that is traceable to the therapy level standard at the National Physical Laboratory is presented. Correction factors have been generated to account for the finite source and detector size . The air kerma rate calibration of the secondary standard for caesium quality has been derived by polynomial curve fitting. The reference a ir kerma rates for several caesium tubes and needles have been determi ned and the results compared with the manufacturers' source test repor ts. For all the source types used the agreement between methods was wi thin 2%.