Na. Deshpande et Jm. Wilkinson, CALIBRATION OF LOW ACTIVITY CESIUM TUBES AND NEEDLES TRACEABLE TO THETHERAPY LEVEL STANDARD, British journal of radiology, 67(794), 1994, pp. 194-199
A technique for deriving the reference air kerma rate for low activity
brachytherapy sources that is traceable to the therapy level standard
at the National Physical Laboratory is presented. Correction factors
have been generated to account for the finite source and detector size
. The air kerma rate calibration of the secondary standard for caesium
quality has been derived by polynomial curve fitting. The reference a
ir kerma rates for several caesium tubes and needles have been determi
ned and the results compared with the manufacturers' source test repor
ts. For all the source types used the agreement between methods was wi
thin 2%.