Jlm. Venselaar et al., INTERCOMPARISON OF CALIBRATION PROCEDURES FOR IR-192 HDR SOURCES IN THE NETHERLANDS AND BELGIUM, Radiotherapy and oncology, 30(2), 1994, pp. 155-161
Citations number
12
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Thirteen radiotherapy departments in The Netherlands and Belgium parti
cipated in an intercomparison of the calibration procedures in use for
the quality control of iridium-192 high dose rate sources. A measurin
g device from 1 of the departments was taken to the 12 other participa
nts, where 1-3 sources were measured by both the local and the visitin
g physicist's team. Also, 1 Ir-192 HDR reference source was taken to 7
other - Dutch - departments to serve as a stability check of the meas
uring device. the results of the local calibration were compared with
the values obtained by the visiting team. A mean deviation of + 1.3% w
as found in 19 observations, with values ranging from -0.4% to +3.0%.
In total, 16 sources were checked. The measured reference air kerma ra
te of the available sources was compared with the value stated on the
source calibration certificate: the values on the certificate were on
the average +1.8% higher than the measured values, ranging from -4.2%
to +6.8%. In 1 instance the reference source was also measured by The
Netherlands Measurements Institute (NMI), the national standards labor
atory in The Netherlands. A good agreement was found between the NMI c
alibration and the reference air kerma rate determined with the measur
ing device used in this project.