C. Weltens et al., RELIABILITY OF CLINICAL PORT FILMS FOR MEASURING DOSE INHOMOGENEITIESIN RADIOTHERAPY FOR HEAD AND NECK TUMORS, Radiotherapy and oncology, 30(2), 1994, pp. 167-170
Citations number
10
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Reproducibility and reliability of film dosimetry are evaluated in cli
nical circumstances. Optical densities were read on 82 routinely taken
verification films of Ear-Nose-Throat (ENT) treatment fields and corr
ected with the sensitometric curve to obtain relative dose values. Off
-axis ratios measured on 10 successive films of one specific field wer
e compared with each other and show a good reproducibility (s = 1.7%).
Reliability was assessed by comparing 82 off-axis ratios obtained on
the films with off-axis ratios obtained by simultaneously performed di
ode measurements. Only small relative differences were detected (xBAR
= 1.1%, s = 3.0).