NEW STROBOSCOPIC LIGHT-SOURCE FOR INTRAOPERATIVE RETINAL FLUORESCEIN ANGIOGRAPHY

Citation
Rr. Krueger et al., NEW STROBOSCOPIC LIGHT-SOURCE FOR INTRAOPERATIVE RETINAL FLUORESCEIN ANGIOGRAPHY, Archives of ophthalmology, 112(3), 1994, pp. 420-422
Citations number
11
Categorie Soggetti
Ophthalmology
Journal title
ISSN journal
00039950
Volume
112
Issue
3
Year of publication
1994
Pages
420 - 422
Database
ISI
SICI code
0003-9950(1994)112:3<420:NSLFIR>2.0.ZU;2-M
Abstract
We tested a new stroboscopic light source for intraoperative fluoresce in angiography. A pulsed xenon light source and narrow-band interferen ce filters are coupled to a 20-gauge endoilluminator and a 35-mm camer a mounted on a surgical microscope. The system was tested with good su ccess in a patient undergoing penetrating keratoplasty. An 18- to 22-m uJ flash of 2 milliseconds' duration is achieved, which is well below the safety threshold level. This new stroboscopic light source provide s a flash of sufficient light intensity to perform high-resolution, hi gh-contrast fluorescein angiography during operations for the removal of opacified media.