T. Hayashi et al., DOSE-RATE DEPENDENCE OF CELLULOSE TRIACETATE DOSIMETER AND RADIOCHROMIC FILM DOSIMETER, Radiation physics and chemistry, 43(6), 1994, pp. 613-616
Citations number
17
Categorie Soggetti
Nuclear Sciences & Tecnology","Chemistry Physical","Physics, Atomic, Molecular & Chemical
The doses of gamma-rays and electron beams were evaluated by Fricke do
simetry and the responses of cellulose triacetate dosimeter (CTA) and
radiochromic film dosimeter (RCF) to the two types of radiations were
compared to investigate their dose rate dependence. Both the change in
absorbance at 280 nm of CTA and that in absorbance at 510 nm of RCF c
aused by gamma-irradiation were larger than those by electron-irradiat
ion, when the dosimeters were irradiated to the same dose. The results
in this study suggest that the responses of CTA and RCF are dependent
upon dose rate.