CONVERSION OF THE FINNIGAN-MAT TSQ-70 THERMOSPRAY IONIZATION INTERFACE TO AN ELECTROSPRAY-IONIZATION INTERFACE

Authors
Citation
S. Jackett et M. Moini, CONVERSION OF THE FINNIGAN-MAT TSQ-70 THERMOSPRAY IONIZATION INTERFACE TO AN ELECTROSPRAY-IONIZATION INTERFACE, Review of scientific instruments, 65(3), 1994, pp. 591-596
Citations number
21
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
65
Issue
3
Year of publication
1994
Pages
591 - 596
Database
ISI
SICI code
0034-6748(1994)65:3<591:COTFTT>2.0.ZU;2-U
Abstract
The conversion of a thermospray ionization (TSI) feature of a Finnigan -MAT TSQ-70 mass spectrometer to electrospray ionization (ESI) and its performance are described. The existing source, pumping capacity, fla nges, and temperature controller of the TSQ-70 TS feature were used wi th a few modifications. Conversion of the commercial TS option to a si mple and economically viable ES option has made the analysis of large biomolecules possible without expensive upgrades. To preserve the simp licity of the conversion, desolvation is effected by a heated-capillar y tube (HCT). The HCT and its housing are inserted inside the TSQ-70 T S flange like a solid probe. Mass spectrometric results of low and hig h molecular weight biomolecules, the mass accuracy, sensitivity, and c harge states of the observed ions are comparable to published results by other laboratories. Adequate spectral quality was obtained at short scan times, a required characteristic for interfacing ESI with separa tion methods such as capillary zone electrophoresis.