S. Jackett et M. Moini, CONVERSION OF THE FINNIGAN-MAT TSQ-70 THERMOSPRAY IONIZATION INTERFACE TO AN ELECTROSPRAY-IONIZATION INTERFACE, Review of scientific instruments, 65(3), 1994, pp. 591-596
The conversion of a thermospray ionization (TSI) feature of a Finnigan
-MAT TSQ-70 mass spectrometer to electrospray ionization (ESI) and its
performance are described. The existing source, pumping capacity, fla
nges, and temperature controller of the TSQ-70 TS feature were used wi
th a few modifications. Conversion of the commercial TS option to a si
mple and economically viable ES option has made the analysis of large
biomolecules possible without expensive upgrades. To preserve the simp
licity of the conversion, desolvation is effected by a heated-capillar
y tube (HCT). The HCT and its housing are inserted inside the TSQ-70 T
S flange like a solid probe. Mass spectrometric results of low and hig
h molecular weight biomolecules, the mass accuracy, sensitivity, and c
harge states of the observed ions are comparable to published results
by other laboratories. Adequate spectral quality was obtained at short
scan times, a required characteristic for interfacing ESI with separa
tion methods such as capillary zone electrophoresis.