Ma. Bernstein et Pe. Licato, ANGLE-DEPENDENT UTILIZATION OF GRADIENT HARDWARE FOR OBLIQUE MR-IMAGING, Journal of magnetic resonance imaging, 4(1), 1994, pp. 105-108
A straightforward method for calculating maximum gradient amplitudes f
or the frequency-encoding, phase-encoding, and section-select function
s in oblique magnetic resonance (MR) imaging is presented. The calcula
ted amplitude values depend on the prescribed angulation and on which
of the three functions arc simultaneously active. The derived expressi
ons are optimal in most cases, and when not optimal, they still offer
substantial improvement over calculations that do not incorporate angu
lar information. Both matrix-based and geometric interpretations are p
resented.