N. Kaneki et al., ELECTROCHEMICAL ENZYME-IMMUNOASSAY USING SEQUENTIAL SATURATION TECHNIQUE IN A 20-MUL CAPILLARY - DIGOXIN AS A MODEL ANALYTE, Analytica chimica acta, 287(3), 1994, pp. 253-258
Capillary enzyme immunoassay with flow-injection analysis for digoxin
using the sequential saturation technique has been developed. Glass ca
pillary tubes (10 cm x 0.53 mm i.d.) with immobilized digoxin antibody
were used as the immunoassay reactor. The product of enzymatic reacti
on, 4-aminophenol, was detected amperometrically. The digoxin and the
labeled digoxin binding reaction with the immobilized digoxin antibody
were completed in 2 and 10 min, respectively. Digoxin was determined
in a 20-mu l sample with a detection limit of 10 pg ml(-1) (200 fg or
260 attomoles) and a 3 orders of magnitude range.