ARE THIN-LAYER CHROMATOGRAPHIC SCREENING PROCEDURES STILL UP-TO-DATE - CRITICAL COMPARISON WITH IMMUNOCHEMICAL TESTS

Citation
H. Schutz et al., ARE THIN-LAYER CHROMATOGRAPHIC SCREENING PROCEDURES STILL UP-TO-DATE - CRITICAL COMPARISON WITH IMMUNOCHEMICAL TESTS, Die Pharmazie, 49(2-3), 1994, pp. 213-216
Citations number
29
Categorie Soggetti
Pharmacology & Pharmacy",Chemistry
Journal title
ISSN journal
00317144
Volume
49
Issue
2-3
Year of publication
1994
Pages
213 - 216
Database
ISI
SICI code
0031-7144(1994)49:2-3<213:ATCSPS>2.0.ZU;2-9
Abstract
The formerly prefered TLC screening procedures are often replaced by i mmunoassays. Many pitfalls must be taken into account when enzyme-immu nological (e.g. EMIT-st, ETS) or fluorescence-polarisation-immunologic al (e.g. ADx, TDx) screening tests are applied. False negative results may be caused by conjugate formation, poor cross-reactivities and/or low concentrations. This article describes typical cases and indicates modifications which enable the toxicologist to obtain valid results. Beside this the advantages of TLC-methods are described and discussed.