E. Lo et al., CALIBRATION AND TEST OF THE TANGENTIAL PHASE-CONTRAST IMAGING DIAGNOSTIC ON CDX-U, Review of scientific instruments, 68(2), 1997, pp. 1206-1216
A novel CO, laser-based tangential imaging diagnostic was completely c
alibrated and the system was tested on CDX-U plasmas. It was shown tha
t localized, two-dimensional images of the plasma electron density flu
ctuations in the tokamak core can be obtained from a tangential imagin
g beam [E. Lo, J. Wright, and R. Nazikian, Rev. Sci. Instrum. 66, 1180
(1995)]. A variation of the Zernike phase mirror is used. Test measur
ements made on sound waves verify that the system will image phase var
iations placed at the object plane. An absolute sensitivity of approxi
mate to 1 x 10(7) cm(-3)/root Hz was determined. A series of plasma me
asurements made on CDX-U is described. Results show that the plasma fl
uctuations are two to three orders of magnitude above the noise floor
of the diagnostic. It is also verified that a high-pass cut-off wave n
umber (from approximate to 1 to 8 cm(-1)) for the fluctuations can be
selected by translating the phase mirror. The density fluctuation k sp
ectrum is measured and found to peak at approximate to 1-3 cm(-1). (C)
1997 American Institute of Physics.