A 6-week clinical evaluation of the One Touch 11 blood glucose meter w
as performed in a medical ward specializing in diabetic patients and c
ompared to the existing Glucometer II system. Nursing staff trained in
both methods performed 267 capillary blood glucose measurements using
the two meters.A reference capillary blood glucose was performed simu
ltaneously on 129 of these occasions using a Yellow Springs Instrument
analyser. The One Touch II correlated better with the reference (r =
0.97) than the Glucometer II (r = 0.86) and was equally close or close
r to the Yellow Springs Instrument on 82 % of occasions. A greater pro
portion of samples measured by the One Touch II were within 15% of the
reference value (83 % vs 66 %). There were also 17 cases where the On
e Touch II was able to give a measurement which was outwith the analyt
ical range of the Glucometer II. A questionnaire revealed the nursing
staff unanimously preferred using the One Touch II to their present me
ter. It is concluded that in a ward setting the One Touch II meter sho
wed improved accuracy, usefulness in its wider analytical range, and a
greater user preference in comparison to the Glucometer II.