K. Riahi et al., INFLUENCE OF LASER FOCUSING IN MATRIX-ASSISTED LASER-DESORPTION IONIZATION, Rapid communications in mass spectrometry, 8(3), 1994, pp. 242-247
The influence of laser focusing on the target in matrix-assisted laser
desorption/ionization (MALDI) has been studied. The emission of stabl
e and metastable molecular ions of bovine insulin depends on both the
laser energy and the distance between the target and the laser focal p
lane. The results clearly show that the laser fluence is the important
parameter in MALDI. The fluence threshold of metastable-ion emission
is higher than that of stable-ion emission. Beyond this threshold, the
emission of stable ions reaches a plateau. This is due to the spatial
inhomogeneity of the laser energy distribution. A phenomenological mo
del including two fluence thresholds for the stable- and metastable-io
n emission has been developed and compared to the experimental data.