The measurement of changes in ion activity (e.g., pH) in neurons requi
res fine tip-sized double-barreled microelectrodes: one channel being
equipped with an ion-selective liquid membrane, the other used for mea
surement of the membrane potential. The limited bandwidth and the diff
ering transfer functions for electrical and ionic signals necessitate
frequency response linearization networks to ensure that the output si
gnal of the electrode is a faithful image of the input signal. We have
developed a linearization network to ensure a rapid response time for
ion-sensitive microelectrodes. To test the response characteristic we
have developed a test system that allows the pH at the electrode tip
to be changed within 1 ms. Application of these techniques to electrod
es of 1 mu m tip diameter results in a 90% response time to a pH step
of similar to 60 ms and of similar to 2 ms with electrodes with 20 mu
m tip.