FLOW-TIME ANALYSIS DURING SONOGRAPHIC FLO W MEASUREMENTS - IN-VITRO STUDIES TO TEST THEIR ACCURACY

Citation
H. Riediger et al., FLOW-TIME ANALYSIS DURING SONOGRAPHIC FLO W MEASUREMENTS - IN-VITRO STUDIES TO TEST THEIR ACCURACY, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 160(4), 1994, pp. 344-348
Citations number
10
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
09366652
Volume
160
Issue
4
Year of publication
1994
Pages
344 - 348
Database
ISI
SICI code
0936-6652(1994)160:4<344:FADSFW>2.0.ZU;2-3
Abstract
Flow-time analysis is a new technique for colour coded sonographic flo w measurements. The measurements depend on a pattern of ultrasound pul ses and have some significant advantages compared with Doppler procedu res. The accuracy of the method and possible sources of error have bee n studied under experimental conditions. Measurements on continuous bl ood flow showed a correlation of r > 0.99 for duplex sonography and of r > 0.93 for flow-time analysis. For this method, the colour coding p arameters must be optimal in order to achieve high accuracy. Under the se conditions quantitative measurement of pulsatile flow also becomes possible during colour mode.