FLOW-INJECTION ANALYSIS WITH FOURIER-TRANSFORM INFRARED DETECTION FORCLINICAL AND PROCESS ANALYSIS

Citation
E. Rosenberg et R. Kellner, FLOW-INJECTION ANALYSIS WITH FOURIER-TRANSFORM INFRARED DETECTION FORCLINICAL AND PROCESS ANALYSIS, Fresenius' journal of analytical chemistry, 348(8-9), 1994, pp. 530-532
Citations number
4
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
348
Issue
8-9
Year of publication
1994
Pages
530 - 532
Database
ISI
SICI code
0937-0633(1994)348:8-9<530:FAWFID>2.0.ZU;2-Z
Abstract
As the analysis of glucose and urea is of special interest in the fiel d of clinical and process analysis, an enzymatic FIA-FTIR system is pr esented, based on the reactions of glucose to gluconic acid using gluc ose oxidase as well as of urea to ammonium carbonate catalyzed by urea se. The method is tested with standard solutions, with samples simulat ing the blood composition and with a variety of fruit juices and soft drinks. Whereas the systematic error is within few percent, as conclud ed from the slope of the calibration graph, the reproducibility of the method is up to now not satisfactory for routine applications in proc ess analysis and has to be improved.