A NEW CONCEPT OF A HIGH-SENSITIVITY TISSUE-EQUIVALENT PROPORTIONAL COUNTER FOR INDIVIDUAL NEUTRON DOSIMETRY

Citation
B. Boutruche et al., A NEW CONCEPT OF A HIGH-SENSITIVITY TISSUE-EQUIVALENT PROPORTIONAL COUNTER FOR INDIVIDUAL NEUTRON DOSIMETRY, Radiation protection dosimetry, 52(1-4), 1994, pp. 335-338
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
52
Issue
1-4
Year of publication
1994
Pages
335 - 338
Database
ISI
SICI code
0144-8420(1994)52:1-4<335:ANCOAH>2.0.ZU;2-#
Abstract
This paper deals with a new concept of tissue-equivalent proportional counter for personal radiation protection monitoring. A high sensitivi ty to neutron and gamma radiation is obtained by the use of a multicha nnel cathode. Electrons and ions drift in an electric field created in each channel by a bias current inside the cathode material. Three sta ges are described, a single-channel counter with a needle anode, a one and five channel counter with a wire anode and a multicellular counte r with nine wire anodes and more than 250 channels. The first results on the electric field distribution and electron behaviour are presente d. A quasi Dirac chord length distribution is determined. An example o f the microdosimetric spectrum obtained with a monoenergetic neutron i rradiation is shown.