VARIANCE-COVARIANCE MEASUREMENTS IN THERAPEUTIC AND DIAGNOSTIC RADIATION BEAMS

Citation
Hb. Honore et al., VARIANCE-COVARIANCE MEASUREMENTS IN THERAPEUTIC AND DIAGNOSTIC RADIATION BEAMS, Radiation protection dosimetry, 52(1-4), 1994, pp. 387-390
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
52
Issue
1-4
Year of publication
1994
Pages
387 - 390
Database
ISI
SICI code
0144-8420(1994)52:1-4<387:VMITAD>2.0.ZU;2-T
Abstract
Cylindrical tissue-equivalent twin ionisation chambers have been devel oped based on experience with proportional counters for application in microdosimetric variance-covariance measurements of the dose mean lin eal energy, y(D)BAR, in therapeutic and diagnostic radiation beam qual ities. It is concluded that the method is applicable but the size of t he detector signals has to be optimised very carefully in respect to n oise reduction and the total relative variance. It is proposed that re sults from the proportional mode and the ionisation mode are compared to obtain reliable values for the microdosimetric beam quality charact erisation necessary to assess clinically relevant differences.