A HIGH-INTENSITY SPUTTER NEGATIVE-ION SOURCE WITH MULTITARGET POSITIONS AND TARGET SCANNERS

Authors
Citation
Gt. Du, A HIGH-INTENSITY SPUTTER NEGATIVE-ION SOURCE WITH MULTITARGET POSITIONS AND TARGET SCANNERS, Review of scientific instruments, 65(4), 1994, pp. 1151-1152
Citations number
6
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
65
Issue
4
Year of publication
1994
Part
2
Pages
1151 - 1152
Database
ISI
SICI code
0034-6748(1994)65:4<1151:AHSNSW>2.0.ZU;2-O
Abstract
The source is based on the principle put forward by Middleton, but mod ified significantly to a multifunction version for variant application s. It is equipped with a spherical molybdenum ionizer, an eight-positi on target wheel, and a special Cs vapor nozzle. Due to the special des ign, target changing and scanning are carried out only by means of rot ations, to simplify the mechanism and automatic control procedures. Th e tube for transmitting Cs vapor, as a resistance, is connected in ser ies in the ionizer supply circuit to reach a high enough temperature f or the vapor to pass. Details of the mechanical design, operation mode s, and test results are given.