INTRAARTERIAL DOPPLER FLOW MEASUREMENTS DURING RENAL-ARTERY ENDOVASCULAR STENT DEPLOYMENT - ABOLITION OF RESTING IMPEDANCE TO BLOOD-FLOW

Citation
Mc. Bates et al., INTRAARTERIAL DOPPLER FLOW MEASUREMENTS DURING RENAL-ARTERY ENDOVASCULAR STENT DEPLOYMENT - ABOLITION OF RESTING IMPEDANCE TO BLOOD-FLOW, Journal of interventional cardiology, 7(2), 1994, pp. 199-202
Citations number
NO
Categorie Soggetti
Cardiac & Cardiovascular System
ISSN journal
08964327
Volume
7
Issue
2
Year of publication
1994
Pages
199 - 202
Database
ISI
SICI code
0896-4327(1994)7:2<199:IDFMDR>2.0.ZU;2-2
Abstract
This communication details the first reported case of renal artery int raarterial pulse wave Doppler obtained prior to, and following primary stenting of a renal artery stenosis. Doppler wave forms were recorded proximal, within, and distal to the lesion and reported with the angi ographic and hemodynamic results. The changes in average and maximal p eak velocities following stent deployment continued the abolition of t he resting transstenotic gradient and renal artery stenosis. Intraarte rial pulsed wave Doppler complements angiography and translesional gra dient measurements as a physiological marker of a successful procedure .