ENERGY RESOLUTION OF A LEAD SCINTILLATING-FIBER ELECTROMAGNETIC CALORIMETER

Citation
J. Budagov et al., ENERGY RESOLUTION OF A LEAD SCINTILLATING-FIBER ELECTROMAGNETIC CALORIMETER, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 343(2-3), 1994, pp. 476-483
Citations number
15
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
343
Issue
2-3
Year of publication
1994
Pages
476 - 483
Database
ISI
SICI code
0168-9002(1994)343:2-3<476:EROALS>2.0.ZU;2-A
Abstract
A calorimeter module was fabricated using profiled lead plates and sci ntillating fibers with diameter 1 mm and attenuation length about 80 c m. The absorber-to-fiber volume ratio was 1. 17 and the module average radiation length X0 congruent-to 1.05 cm. The energy resolution of th e module was investigated using the electron beams of U-70 at Serpukha v and of the SPS at CERN in the energy range 5-70 GeV. The energy reso lution at theta = 3-degrees (the angle between the fiber axis and the beam direction) may be expressed by the formula sigma/E (%) = 13.1/squ are-root E + 1.7. The energy resolution was also simulated by Monte Ca rlo and good agreement with the experiment has been achieved.