Dm. Stroop et al., COMPARISON OF THE ABBOTT CELL-DYN-3000-SL AND THE COULTER-STKS HEMATOLOGY ANALYZERS, Annals of clinical and laboratory science, 24(3), 1994, pp. 250-258
A side by side comparison of the Cell Dyn 3000 SL and Coulter STKS usi
ng identical samples was performed over a four month period in the Cli
nical Hematology Laboratory at the University of Cincinnati Medical Ce
nter. A total of 444 samples comprised of 114 sophomore medical studen
ts and 330 randomly selected clinical patients were compared in 20 hem
ogram parameters, and the ability of each instrument to generate diffe
rential suspect flags was analyzed. Correlation coefficients for leuko
cytes, erythrocytes, hemoglobin, hematocrit, and platelets were 0.99.
Correlation coefficients for the Cell Dyn and STKS compared with a 600
cell manual differential were 0.98 and 0.89 for neutrophils, 0.96 and
0.87 for lymphocytes, and 0.72 and 0.48 for monocytes, respectively.
Both instruments demonstrated high precision and accuracy by internal
and external quality control standards. Each analyzer exhibited streng
th as a screening instrument for abnormal cell populations. The Coulte
r STKS had overall sensitivity of 75%, specificity of 94%, 25% false n
egatives, and 6% false positives. Sensitivity, specificity, false nega
tive and positive rates for the Cell Dyn 3000 SL were 68%, 92%, 32%, a
nd 8%, respectively. Based upon this extensive side by side comparison
using an identical sample population, it has been concluded that alth
ough statistically significant systematic bias (p < 0.05) exists betwe
en the two instruments, both analyzers can adequately support the need
s of the clinical hematology laboratory.