Re. Maser et al., USE OF ARTERIAL BLOOD WITH BEDSIDE GLUCOSE REFLECTANCE METERS IN AN INTENSIVE-CARE UNIT - ARE THEY ACCURATE, Critical care medicine, 22(4), 1994, pp. 595-599
Objective: To compare blood glucose values obtained from two different
sampling sites (arterial catheter and capillary from finger puncture)
, which were analyzed by a bedside reflectance meter. A sample was als
o analyzed by standard methods (oxygen electrode oxidation in the labo
ratory). Design: Prospective, cross-sectional clinical study. Setting:
Cardiovascular intensive care unit (ICU) designed for postoperative o
pen-heart surgery patients in a 1,100-bed medical center. Patients: Se
quential sample of 50 patients immediately after open-heart surgery. M
easurements and Main Results: The blood glucose concentration of each
patient was analyzed on the patient's arrival to the ICU (immediately
postoperatively) by three methods: one blood specimen was obtained fro
m an arterial catheter, divided and analyzed either at the bedside by
a reflectance meter (glucose method 1) or in the hospital laboratory (
glucose method 2); another sample was obtained by lancing the fingerti
p and the glucose concentration was analyzed at the bedside in the sam
e reflectance meter (glucose method 3). Using paired analyses to compa
re the mean glucose values of the bedside arterial whole blood sample
(method 1) with the arterial, serum sample (method 2) demonstrated tha
t the glucose concentration in the arterial whole blood sample (method
1) was significantly (p <.001) higher. For 46 of 50 comparisons, the
glucose value in the arterial whole blood sample (method 1) was higher
, with a mean difference of 30 mg/dL (1.7 mmol/L). Although the mean d
ifference was reduced to 10 mg/dL, (0.6 mmol/L) when the arterial whol
e blood sample (method 1) was corrected for the hematocrit (i.e., <35%
[<0.35]), the mean glucose concentration in the arterial whole blood
samples (method 1) remained statistically higher (p <.05). The glucose
concentration in the arterial serum sample (method 2) was significant
ly higher than the value determined from the bedside capillary sample
(method 3) before (p <.05) and after (p <.001) correction for hematocr
it. The difference in mean glucose concentrations between the arterial
serum sample (method 2) and bedside capillary sample (method 3) was 9
mg/dL (0.5 mmo/L) when the capillary specimen (method 3) was not corr
ected for hematocrit. This difference increased to 21 mg/dL (1.2 mmol/
L) when low hematocrit values were considered and appropriate adjustme
nts of the glucose values were made. At the bedside, one can accuratel
y correct arterial whole blood glucose values to correspond to laborat
ory values by the following formula: (0.94 x arterial whole blood gluc
ose) + (4.6 x hematocrit) + (-16.5 x [37 degrees C patient's temperatu
re]) - 132 = laboratory glucose value. Conclusions: Since arterial who
le blood samples give higher glucose results than arterial serum, the
use of arterial whole blood in combination with reflectance meters mus
t be recommended with caution. This caution is especially advised if t
he glucose values obtained with arterial whole blood are used in conju
nction with a sliding scale of insuIin, which depends on threshold con
centrations of glucose. In our hospital, use of arterial whole blood i
n combination with reflectance meters could have resulted in an incorr
ect dose of insulin in 31 of 50 patients.