MONITORING NEUROMUSCULAR BLOCK BY ACCELEROMYOGRAPHY - COMPARISON OF THE MINI-ACCELOGRAPH WITH THE MYOGRAPH-2000

Citation
Njn. Harper et al., MONITORING NEUROMUSCULAR BLOCK BY ACCELEROMYOGRAPHY - COMPARISON OF THE MINI-ACCELOGRAPH WITH THE MYOGRAPH-2000, British Journal of Anaesthesia, 72(4), 1994, pp. 411-414
Citations number
5
Categorie Soggetti
Anesthesiology
ISSN journal
00070912
Volume
72
Issue
4
Year of publication
1994
Pages
411 - 414
Database
ISI
SICI code
0007-0912(1994)72:4<411:MNBBA->2.0.ZU;2-Q
Abstract
The precision of the compact Mini-Accelograph (M-A) was compared with the Myograph 2000 (MYO). Neuromuscular block resulting from atracurium was measured simultaneously by the MYO and the M-A applied to contral ateral thumbs. During onset, the M-A frequently underestimated the ext ent of block (maximal at approximately 50% twitch depression). The M-A control train-of-four (TOF) ratio was characteristically > 1.0 and re mained greater than the MYO ratio during the onset of atracurium. Duri ng recovery, the difference between the MYO and the M-A was maximal at approximately 50% twitch depression, but the M-A frequently overestim ated the extent of block. The mean differences between the MYO and the M-A were small in respect of the recovery index (RI) and the TOF. How ever, the limits of agreement were unacceptably wide for both TOF and RI. When the MYO TOF was 0.7, the corresponding M-A TOF varied between 0.4 and 1.0. The M-A was more susceptible to drift than the MYO.