A NOVEL EDDY-CURRENT COMPENSATION SCHEME FOR PULSED GRADIENT SYSTEMS

Citation
Re. Wysong et al., A NOVEL EDDY-CURRENT COMPENSATION SCHEME FOR PULSED GRADIENT SYSTEMS, Magnetic resonance in medicine, 31(5), 1994, pp. 572-575
Citations number
6
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
07403194
Volume
31
Issue
5
Year of publication
1994
Pages
572 - 575
Database
ISI
SICI code
0740-3194(1994)31:5<572:ANECSF>2.0.ZU;2-1
Abstract
We describe a modified pre-emphasis network that uses a number of fixe d time constants, only whose amplitudes are adjustable. In comparison with variable time constant preemphasis networks we have used, the fix ed time constant network is more effective and easier to set on the fl y. We have further developed a method to set the fixed time constant n etwork mathematically using a linear least squares technique.