AUTOMATED, HIGH-PRECISION COULOMETRIC TITRIMETRY .2. STRONG AND WEAK ACIDS AND BASES

Authors
Citation
Kw. Pratt, AUTOMATED, HIGH-PRECISION COULOMETRIC TITRIMETRY .2. STRONG AND WEAK ACIDS AND BASES, Analytica chimica acta, 289(2), 1994, pp. 135-142
Citations number
13
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
289
Issue
2
Year of publication
1994
Pages
135 - 142
Database
ISI
SICI code
0003-2670(1994)289:2<135:AHCT.S>2.0.ZU;2-W
Abstract
Automated constant-current coulometric acidimetry, based on Faraday's Laws, is uncertain to less than 1 part in 20000 (relative standard dev iation) and requires no chemical standardization. It is applicable to strong and weak acids and bases, with bases back-titrated after additi on of excess strong acid. Initial setup and sample introduction are th e sole manual steps. Assays of HCl, benzoic acid, Na2CO3, and tris(hyd roxymethyl)aminomethane are presented. In the endpoint determination p rocedure, a generalized titration equation yields the theoretical char ge remaining to the endpoint before each charge addition. The ratio of the experimental to the theoretical charge for the preceding aliquot corrects for experimental deviations.