NOVEL MERCURY-COATED CARBON-FIBER VOLTAMMETRIC DETECTOR FOR USE IN ADSORPTIVE STRIPPING FLOW-ANALYSIS

Citation
Ja. Delpozo et al., NOVEL MERCURY-COATED CARBON-FIBER VOLTAMMETRIC DETECTOR FOR USE IN ADSORPTIVE STRIPPING FLOW-ANALYSIS, Analytica chimica acta, 289(2), 1994, pp. 169-176
Citations number
26
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
289
Issue
2
Year of publication
1994
Pages
169 - 176
Database
ISI
SICI code
0003-2670(1994)289:2<169:NMCVDF>2.0.ZU;2-A
Abstract
The preparation and application of a novel mercury-coated carbon fibre flow microelectrode are reported. The possibility to use this device in a flow system combined with alternating current (ac) stripping volt ammetry is also demonstrated using mitoxantrone (MXT) as probe compoun d. The electrode is easy to prepare. Special analytical advantages hav e been observed when a.c. stripping voltammetry is applied to the adso rbed molecule on the electrode due to the reversibility of the electro chemical process. Once the operating system has been optimized a limit of detection for mitoxantrone of 4.5 x 10(-11) M was found. Manipulat ion of flow conditions allows the study of different concentration ran ges. The direct analysis of serum samples without sample pretreatment, is also discussed and provided a linear calibration range between 4.5 X 10(-7) M and 4.5 x 10(-6) M.