Ed. Demuinck et al., MYOCARDIAL PROTECTION DURING CORONARY ANGIOPLASTY WITH AUTOPERFUSION AND FORCED PERFUSION - AN IN-VITRO COMPARISON, International journal of artificial organs, 17(2), 1994, pp. 83-87
During coronary angioplasty, perfusion distal to the inflated angiopla
sty balloon can be maintained with autoperfusion balloon catheters and
coronary perfusion pumps. The blood flow rates through the autoperfus
ion balloon catheters and the flow rates achieved with a perfusion pum
p were compared in vitro with fresh human blood at 37 degrees C. In a
specially designed system, blood flow rates through StackTM autoperfus
ion balloon catheters were measured at 40, 60 and 80 mmHg continuous p
ressure. In another system, driving pressures were measured during per
fusion with the pump, through a specially designed forced perfusion ca
theter at 20, 40 and 60 ml/min flow. The pressure applied in the autop
erfusion experiments was converted into atmospheres (atm) to facilitat
e comparison with the driving pressures measured during pumping (1 mmH
g = 1.316 x 10(-3) atm). Mean flow rates through the autoperfusion bal
loon catheters were: 46 ml/min at 0.05 atm, 66 ml/min at 0.09 atm and
75 ml/min at 0.1 atm. Mean pressures during pumping were: 1.8 atm at 2
0 ml/min, 3.5 atm at 40 ml/min, 5 afm at 60 ml/min. Due to the phasic
nature of coronary blood flow, the flow through autoperfusion balloons
is generally lower than the minimum required for adequate myocardial
protection (= 60 ml/min). Thus, autoperfusion balloon catheters are si
mpler and cheaper devices than perfusion pumps, but generally they are
not able to provide adequate myocardial protection.